Data format UPP (Universal Pyrometer Protocol) 34

Description

Command

Parameters

External clearing: AAlx

Clearing the max. value storage (only with lz = 7, external)

Analog output:

AAasX

X = 0 or 1

0 = 0 … 20 mA;1 = 4 … 20 mA

Reading basic
temperature
range:

AAmb

Output: XXXXYYYY (hex 8-digit, °C or °F)

XXXX = beginning of temp. range
YYYY = end of temp. range

Reading
temperature sub
range:

AAme

Output: XXXXYYYY (hex 8-digit, °C or °F)

XXXX = beginning of temp. range (°C or °F)
YYYY = end of temp. range (°C or °F)

Setting of

temperature sub
range:

AAm1XXXXYYYY XXXX (hex 4-digit) beginning of temp. range

YYYY (hex 4-digit) end of temp. range

Address:

AAgaXX

XX = (00 ... 97)

00 ... 97 = regular device addresses
99 = Global address with response
98 = Global address without response
(only setting commands!)

Baud rate:

AAbrX

X=0...5 (decimal)

0 = 1200 Baud

2 = 4800 Baud 4 = 19200 Baud

1 = 2400 Baud

3 = 9600 Baud

5 = 38400 Baud

Temp. display
°C / °F

AAfhX

Output: X = 0: display in °C; X = 1: display in °F

Wait time:

AAtwXX

XX = 00 ... 99 (decimal, in bit time of the current baud
rate)

Internal

temperature:

AAgt

Output: XXX (dec. 000 ... 099°C, 032 ... 210°F)

Max. internal
temperature:

AAtm

Output: XXX (dec. 000 ... 099, always in °C)

Limit switch SP1:

AAslXXXX

XXXX = set Limit switch, hex 4-digit, °C or °F

Mode
Limit switch SP1:

AAt1X

X = 0 Limit switch Off
X = 1 Limit switch close above
X = 2 Limit switch close below

Hysteresis
Limit switch SP1:

AAhlXX

XX = set hysteresis, hex 2-digit, °C or °F

Error status:

AAfs

Output:

XX;

XX=00…FF

(00

=

no

error)

(01…FF: error code for LumaSense service)

LED aiming light: AAlaX

X = 0 switch off aiming light
X = 1 switch on aiming light

LED aiming light
at Power-on

AAlpX

X = 0 aiming light off at Power-on
X = 1 aiming light on at Power-on

Reading
parameters:

AApa

Output 11-digit, decimal:

Digit 1 und 2 (10...99 or 00): Emissivity
Digit 3 (0 ... 6): Exposure time

Digit 4 (0 ... 8): Clear time max. / min. value storage
Digit 5 (0 ... 1): Analog output
Digit 6 und 7: (00 ... 98): Internal temperature
Digit 8 und 9 (00 ... 97): Address
Digit 10 (0 ... 6 or 8): Baud rate
Digit 11 always 0

Device type:

AAna

Output: “IGA 320/23-LO” (16 ASCII-characters)

Serial number:

AAsn

Output: XXXXX (dec. 5-digit)