IGA 320/23-LO Operating Manual

Data format UPP (Universal Pyrometer Protocol) 35

Description

Command

Parameters

Device type /

software version:

AAve

Output: XXYYZZ (6-digit decimal)

XX = 56 (IGA 320/23)
YY = Month of software version
ZZ = Year of software version

Detailed
Software
version:

AAvs

tt.mm.yy XX.YY
tt = day; mm = month; yy = year; XX.YY = software version

Reference
number:

AAbn

Output: XXXXXX (hex 6-digit)

Note: the letter “l” means the lower case letter of “L”.

Additional instructions for the RS485 interface:

Requirements to the master system during half-duplex operation:

1. After an inquiry, the bus should be switched into a transmission time of 3 bits

(some older interfaces are not fast enough for this).

2. The pyrometer's response will follow after 5 ms at latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be

repeated.