10 troubleshooting – LumaSense Technologies IGA 320/23-LO User Manual

Page 41

Advertising
background image

IGA 320/23-LO Operating Manual

Troubleshooting 41

10 Troubleshooting

Before sending the pyrometer for repair, try to find the error and to solve the problem with the

help of the following list.

Temperature indication too low

Incorrect alignment of the pyrometer to the object

New correct alignment to achieve the max. temperature signal

Measuring object smaller than spot size

check measuring distance, smallest spot size is at nominal measuring distance

Measuring object is not always in the measuring spot of the pyrometer

Use max. value storage (see

4.9

)

Emissivity set too high

Set lower correct emissivity corresponding to the material (see

4.5

)

Lens contaminated

Clean lens carefully (see

6.2

)

Temperature indication too high

Emissivity set too low

Set lower correct emissivity corresponding to the material (see

4.5

)

The measurement is influenced by reflections of hot machine parts

Use mechanical construction to avoid the influence of the interfering radiation

Measuring errors

Indicated temperature is decreasing during the use of the pyrometer, contamination of the

lens

Clean lens. Recommendation: use of air purge (see

3.3

)

Air contamination in the sighting path between pyrometer and object

Change position of the pyrometer with a clean sighting path (if necessary use a ratio

pyrometer)
HF-interferences

Correct the connection of the cable shield (see

3.1

)

Temperature Indication is fluctuating, probably caused by changing emissivity

Wrong pyrometer type, use of ratio pyrometer recommended