LumaSense Technologies INNOVA 1309 User Manual

Page 28

Advertising
background image

Chapter 3
________________________________________________________________________

________________________________________________________________________

BE1285-15

1309 Multipoint Sampler

LumaSense Technologies A/S

Page 28 of 42

Table 3.2. The 1309’s Status Flag. Each bit represents a particular

1309 component; when set, the 1309 is currently using that
component.

Status Flag

Current function when set

Bit

no.

Dec.

value

Bit name

1

1 Sampling valve 1

Sampling valve 1 open

2

2 Sampling valve 2

Sampling valve 2 open

3

4 Sampling valve 3

Sampling valve 3 open

4

8 Sampling valve 4

Sampling valve 4 open

5

16 Sampling valve 5

Sampling valve 5 open

6

32 Sampling valve 6

Sampling valve 6 open

7

64 Sampling valve 7

Sampling valve 7 open

8

128 Sampling valve 8

Sampling valve 8 open

9

256 Sampling valve 9

Sampling valve 9 open

10

512 Sampling valve 10 Sampling valve 10 open

11

1024 Sampling valve 11 Sampling valve 11 open

12

2048 Sampling valve 12 Sampling valve 12 open

13

4096 3-way valve

3-way valve set to analyzer

14

8192 Not used

Not used

15

16384 Not used

Not used

16

32768 Not used

Not used

Example:

STATUS?<LF>

Outputs the Status flag of the 1309

as described above. Refer to Table
3.2 for the decimal values of each

bit. Refer also to the examples giv-
en below.

If the 1309 outputs 4352, using Table 3.2 shows that 4352 = 256 +

4096. Therefore, bits 9 and 13 are set. The mechanical condition of the
1309 is: sampling valve number 9 open, and 3-way valve set to Ana-

lyzer.

3.6.2

Reset System Interface Job

This job restarts the 1309. The 1309 restarts as described in

section

3.1

. Resetting the 1309 cancels any task which the 1309 was perform-

ing prior to the reset.

Example:

RESET_SYSTEM<LF>

Resets the 1309.