6 technical specifications – HT instruments HT4010 User Manual

Page 36

Advertising
background image

HT4010

EN - 15

6 TECHNICAL

SPECIFICATIONS

6.1 CHARACTERISTICS
Accuracy is indicated as [% rdg + dgt]. It is referred to: 23°C ± 5°C with RH <80%RH

DC Voltage

Range Resolution Accuracy

Input

impedance

Overload

protection

200.0mV 0.1mV

±(0.8%rdg + 2dgt)

10M

600VDC/ACrms

2.000V 0.001V

±(1.5%rdg + 2dgt)

20.00V 0.01V
200.0V 0.1V

600V 1V

±(2.0%rdg + 2dgt)


AC Voltage

Range Resolution

Accuracy

(50

÷ 60Hz)

Input

impedance

Overload

protection

200.0mV 0.1mV

±(1.5%rdg + 3.5mV)

10M

600VDC/ACrms

2.000V 0.001V

±(1.8%rdg + 8dgt)

20.00V 0.01V
200.0V 0.1V

600V 1V

±(2.5%rdg + 8dgt)


AC Current

Range Resolution Accuracy

Frequency

range

Overload

protection

2.000A 0.001A

±(2.5%rdg + 10dgt)

50

÷60Hz

600Arms

20.00A 0.01A

±(2.5%rdg + 4dgt)

200.0A 0.1A

600A 1A

±(4.0%rdg + 8dgt)

Resistance and Continuity test

Range Resolution Accuracy

Buzzer

Overload

protection

200.0

Ω 0.1Ω

±(1.0%rdg + 4dgt)

≤150

600VDC/ACrms

2.000k

0.001k

±(1.5%rdg + 2dgt)

20.00k

0.01k

200.0k

0.1k

2.000M

0.001M

±(2.5%rdg + 3dgt)

20.00M

0.01M

±(3.5%rdg + 5dgt)


Diode test

Features

Test current

Open voltage

0.3mA typical

1.5VDC