Caution – HT instruments HT9015 User Manual

Page 17

Advertising
background image

HT9014 - HT9015

EN - 16

4.3.9. Frequency and Duty cycle measurement

CAUTION

• On frequency test with test leads the maximum input for AC Voltage

measurements is 1000Vrms. Do not take any voltage measurement
exceeding this limit in order not to risk electrical shock or damage the
tester

• On frequency test with jaws make sure that all the test leads are

disconnected from the meter terminals for current measurement.

Fig. 10: Taking frequency and duty cycle measurements


1. Rotate the switch on

position for frequency measurement with test leads or 60A

or 600A positions (only HT9015) or 60A

or 600A∼ positions (only HT9014) for

frequency measurements with jaws

2. Pushing Hz% key the “Hz” symbol for frequency measurement or the “%” symbol for

duty cycle measurement are shown at display

3. Insert the red test lead plug into V

CAPHz%Temp

jack and the black test lead

plug into COM jack (see Fig. 10 – left side) for frequency measurement with test leads.
Open the clamp and put the conductor to be tested in the middle of the clamp jaw (see
Fig. 10 – right side) for frequency measurements with jaws, then reading will be
displayed. Bargraph is disabled in frequency and duty cycle test

4. When “O.L.” symbol is displayed, the result exceeds the maximum value that the

instrument is capable of measuring

5. For HOLD feature please refer to § 4.2