Technical specifications, Caution – HT instruments M71 User Manual

Page 21

Advertising
background image

M71

EN – 20

6. TECHNICAL SPECIFICATIONS

6.1. TECHNICAL

FEATURES

Accuracy is indicated as [% rdg + (dgt number) * resolution] at 23°C ± 5°C,< 80%HR

CAUTION

Under modes EARTH-2W and EARTH-3W the instrument displays the
symbol of warning

whenever:

• It is working under critical conditions, such as in the presence of input

voltages

• It cannot grant the uncertainty of measurement lower than 30% of

reading, according to IEC/EN61557-1


Earth Resistance measurement

Range Resolution

Accuracy

Disturbance

3Vrms

3Vrms<Disturbance<6Vrms

AUTO mode triggered

0.01

÷ 19.99Ω

0.01

±(2%rdg + 3dgt)

±(4%rdg + 10dgt)

20.0

÷ 199.9Ω

0.1

200

÷ 1999Ω

1

2.00

÷ 19.99kΩ

0.01k

20.0

÷ 49.9kΩ

0.1k

±(3rdg + 3dgt)

±(6%rdg + 10dgt)

Measurement frequency

110Hz ± 1Hz

Test current

≤12mA~

Open circuit voltage

≤25Vrms

Resistance values of the voltmetric (Rp) and ampermetric probes (Rc) always meet the following conditions:

¾

If R

A

≤ 10Ω Æ Rc, Rp ≤ 1kΩ

¾

If 10

Ω < R

A

≤ 500Ω Æ Rc, Rp ≤ 100 * R

A

¾

If R

A

> 500

Ω Æ Rc, Rp ≤ 50kΩ


Disturbance Voltage measurement

Range (V)

Resolution (V)

Accuracy

0.0

÷ 99.9

0.1

±(2%rdg + 2dgt)

100

ч 299

1

Advertising