Caution – HT instruments M71 User Manual

Page 8

Advertising
background image

M71

EN – 7

4.1.1. Switching on

When switching on the instrument a brief tone is audible along
with display of all segments for about one second.
Subsequently the last firmware version as well as the last
selected measuring mode are displayed before switching off.


4.1.2. Auto Power OFF
To save battery life the instrument automatically turns off 5 minutes after last pressure of
keys. To resume operation turn on the instrument pressing the start key.

4.2. AUTO

MODE

Starting up this mode the instrument will alter the measuring time to fit to the testing
conditions detected.

CAUTION

• If the instrument detects an “electrical noise” lower than 0.5Vrms before

testing, it will adopt the “standard” test duration (approx. 13 s)

• If the instrument detects an “electrical noise” higher than 0.5Vrms before

testing, it will extend the test duration (approx. 25s) to increase the
outcome stability even under severe measuring conditions