Technical specifications – HT instruments MPP300 User Manual

Page 24

Advertising
background image

MPP300

EN - 22

8. TECHNICAL SPECIFICATIONS

8.1.

TECHNICAL SPECIFICATIONS FOR TEST ON PV SYSTEMS

Uncertainty is indicated as [%reading + (no. of digits) * resolution] at 23°C ± 5°C, <80%HR

DC voltage

Range [V]

Resolution [V]

Uncertainty

10.0

 999.9

0.1

(0.5rdg + 2dgt)

Phase-Neutral AC TRMS voltage

Range [V]

Frequency

Resolution [V]

Uncertainty

10.0

 300.0

42.5

 69.0Hz

0.1

(0.5rdg + 2ddgt)

Max peak factor: 1,5

Phase-Phase AC TRMS voltage

Range [V]

Frequency

Resolution [V]

Uncertainty

50.0

 594.0

42.5

 69.0Hz

0.1

(0.7rdg + 2dgt)

Max peak factor: 1,5

DC current (by means of external clamp transducer)

Range Resolution

Uncertainty

Protection against

overcharge

5.0mV

 319.9mV

0.1mV

(0.5rdg + 0.06%FS)

10V

320.0mV

 999.9mV

(0.5rdg)

Current values corresponding to a voltage < 5mV are zeroed

AC TRMS current (by means of external clamp transducer STD type)

Range Frequency

Resolution Uncertainty

Protection against

overcharge

5.0mV

 219.9mV

42.5

 69.0Hz

0.1mV

(0.5rdg + 0.06%FS)

10V

220.0mV

 999.9mV

(0.5rdg)

Peak factor <= 1.5 – Current values corresponding to a voltage < 5mV are zeroed.

AC TRMS current (by means of external clamp transducer of FLEX 8.5uV/A – FS 100A type)

Range Frequency

Resolution Uncertainty

Protection against

overcharge

0.008

 8.50mV

42.5

 69.0Hz

0.001mV

 (0.5%rdg + 7dgt)

10V

Peak factor <= 1.5 – Current values < 1A are zeroed.

AC TRMS current (by means of external clamp transducer of FLEX 8.5uV/A – FS 1000A type)

Range Frequency

Resolution Uncertainty

Protection against

overcharge

0.085

 85.0mV

42.5

 69.0Hz

0.01mV

 (0.5%rdg + 15dgt)

10V

Peak factor <= 1.5 – Current values < 5A are zeroed.

DC power (Vmeas > 150V) ; AC power (Vmeas > 200V, PF=1)

Clamp FS [A]

Range [W]

Resolution [W]

Uncertainty

1< FS

 10

0.000k

 9.999k

10.00k

 99.99k

0.001k

0.01k

(0.7rdg + 3dgt)

(Imeas < 10%FS)

10< FS

 100

0.00k

 99.99k

100.0k

 999.9k

0.01k

0.1k

(0.7rdg)

(Imeas

 10%FS)

100< FS

 1000

0.0k

 999.9k

1000k

 9999k

0.1k

1k

Vmeas = voltage at which power is measured; Imeas = measured current

Advertising