Heading1 - 5.3 current loop outputs, 3 current loop outputs – In-Situ Con TROLL PRO Installation Manual (purchased before 10 May 2012) User Manual

Page 25

Advertising
background image

Input/Output Connections

25

CONNEXIONS POUR LE RACORD DES SONDES

Danger
Débrancher toute alimentation à l'appareil avant de
connecter les fils.

1. Disconnect power to the instrument.

2. Use a Phillips screwdriver to remove the controller lid.

Note: If you want to use
RuggedCable

®

with

twist-lock connectors,
order a short length of
stripped and tinned
RuggedCable with a
male connector. This
can be wired into the
controller, then
connected to existing
twist-lock cable
connectors.

3. Connect the stripped and tinned wires from the RuggedCable to the

terminal strip as follows:

4. Clip or cap the unused wires (brown and white).

5. Replace the controller lid and tighten screws.

6. Reconnect power to the device.

FIGURE 21. Earth ground

5.3

CURRENT LOOP OUTPUTS

Danger
Make sure that power to the instrument is disconnected
before making any wiring connections.

Danger
The maximum voltage that can be applied across the loop
terminals is 36 V.

SORTIES DE COURANT DE BOUCLE

Danger
Débrancher toute alimentation à l'appareil avant de
connecter les fils.

Danger
La tension maximum à travers la boucle ne doit pas
excéder 36 V.

1. Disconnect power to the instrument.

2. Use a Phillips screwdriver to remove the controller lid.

GND

Black

485A-

Green

485B+

Blue

POWER

Red

Green Earth Ground screw (Figure 21)

Shielded wire (silver color)

Earth ground screws for
sensor silver wires