KANOMAX 2211 Handheld IAQ Monitor User Manual

Page 45

Advertising
background image

40

‘‘‘‘

Use

to select [YES].

Press

to start calibration.


Fully open the regulator valve to pour the gas flow,

Press

to start calibration.











When the calibration result is OK
Calibration is completed.

If [END] is displayed, the valve of regulator will be closed
And gas will be stopped.

Press

to save all datas and return to MENU.

If you press

before you save, you will return to Main Menu

and the setting will not be saved.

Standby screen:

The initial value of countdown is

displayed.

Unit: Second.

………………………………Calibration is performing.

………………………………The remaining time to a calibration end.

……The value after calibration

Operation

…………………Calibration is completed

…………………Calibration result is OK

R

R

e

e

s

s

u

u

l

l

t

t

o

o

f

f

Z

Z

E

E

R

R

O

O

C

C

a

a

l

l

i

i

b

b

r

r

a

a

t

t

i

i

o

o

n

n

:

:

C

C

O

O

c

c

a

a

l

l

i

i

b

b

r

r

a

a

t

t

i

i

o

o

n

n

f

f

a

a

i

i

l

l

s

s

w

w

h

h

e

e

n

n

- CO concentration maintains above

10ppm for more than 30 seconds.

- Value does not fall within 6ppm of

pre-calibration value, which is taken

20 second before calibration.