KANOMAX 2211 Handheld IAQ Monitor User Manual

Page 47

Advertising
background image

42





Use

to select [1.SPAN].

Press

to display a cursor.

Use

to enter a concentration value displayed on a gas

tank, Press

.

Use

to select [YES]. Press





Replace ZERO gas cylinder with SPAN gas tank:
Fully open the regulator valve to pass the gas flow,

press

to start a calibration.

………………Concentration value displayed on gas tank


……The valve before calibration

Operation

Range of CO mass: 20-550ppm
Range of CO

2

mass: 800-5500ppm

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

Standby display

・・・・・・・・・・・・・・・・・・・・・・・・

Shows countdown default value in second.

(CO Calibration:90sec/CO2 Calibration:120sec)

………………………………Calibration is confirming

………………………………The remaining time to a calibration end.