BOP-1K 031014

B.109 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative? QUERY

VOLT:PROT:LIM:NEG?

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:NEG?
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative?
Returns: <exp_value>
Description:Identifies the minimum (maximum absolute) value possible for the negative protection limit
for voltage.

B.110 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive COMMAND

VOLT:PROT:LIM:POS

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:POS <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive <exp_value>
where <value> is between minimum (box) value (Figure 1-3) and 1% above rated (nominal) output
voltage.
Description:Establishes the maximum value possible for the positive protection limit for voltage. This
may be set up to 1% greater than rated voltage. Requires MEM:UPD (PAR. B.12) to save for power
up. See PAR. 3.5.2.2 and Figure 3-16 for special programming considerations.

B.111 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive? QUERY

VOLT:PROT:LIM:POS?

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:POS?
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive?
Returns: <exp_value>
Description:Identifies the maximum value possible for the positive protection limit for voltage.

B.112 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude] COMMAND

VOLT:TRIG

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP] <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Programs voltage value to be transferred to output by *TRG command. Actual output voltage will

depend on load conditions. If the value exceeds the maximum for the model being programmed, error
message -222,”Data out of range” is posted in output queue. If value exceeds VOLT:LIM:HIGH value,
a value corresponding to the voltage limit will be programmed.

B.113 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]? QUERY

VOLT:TRIG?

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: <exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Returns value representing voltage value to be programmed by *TRG command established by

VOLT:TRIG command).

B.114 STATus:OPERation:CONDition? QUERY

STAT:OPER:COND?

Syntax:

Short Form: STAT:OPER:COND?

Long Form: STATus:OPERation:CONDition?

Return Value: <int_value>

Description: Returns the value of the Operation Condition Register (see Table B-3). The Operation Condition

Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-10.)

TABLE B-3. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,

AND OPERATION EVENT REGISTER BITS

CONDITION

NU

LR

NU

LC

SC

CC

TC

CV

NU

TA WTG

NU

BIT

15

14

13

12

11

10

9

8

7

6

5

4 - 0

VALUE

32,768 16,384 8192 4096 2048 1024 512 256 128

64

32

16 -1

CC

Constant Current Mode

CV Constant

Voltage

Mode

LC

List Complete

LR

List Running

NU

Not Used

SC

Sample Complete

TC

Transient Complete

TA

Transient Armed

WTG Waiting For Trigger