FUNCTION GENERATOR CHARACTERISTICS (Continued)

Types of Basic waveforms

Sine (Sinusoidal waveform)

Triangle (Triangular waveform)

Pos. Ramp (Ramp waveform)

Neg. Ramp (Sawtooth waveform)

Square (50% Duty Cycle Pulse)

Level (DC waveform)

Maximum number of points per basic wave-
form

Local: 3933

Count (Number of repetitions)

Range: 1 to 255

For Count = 0 the waveform is exe-

cuted indefinitely until stop com-

mand is sent

• Local: Factory default: when the count is

complete or if execution is stopped using

%

, the output of the unit is set to OFF

(Standby indicator lit). Output must be
enabled (press STANDBY) to execute the
waveform again. (See Table 3-3, WAVE-
FORM STOP for additional settings.)

• Remote: When the count is complete or

the waveform is stopped using the
VOLT:MODE FIX command, the output
remains at last output value.

Waveform Name

10 characters maximum

Waveform Parameters

Type: Voltage or Current

Positive Protection Limit:

(Vd-c or Ad-c)

Negative Protection Limit:

(Vd-c or Ad-c)

Count: number

Initial/Repeating

• Type of waveform, voltage or current, can

be selected only for a new waveform and
consequently applies to all waveform seg-
ments. Changing the type for an existing
waveform is not possible, unless the wave-
form is erased first.

• The Limit values (positive/negative) and

the count number apply to all waveform
segments and can be edited at the begin-
ning of a new waveform or later when mod-
ifying a waveform (see PAR. 3.3.9.7).

• All segments except the last may be set to

run once (Initial) or per the Count setting
(Repeating).

Basic waveform parameters

Sine: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c),

Start Phase (°), Stop Phase (°).

Triangular: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c), Start

Phase (°), Stop Phase (°).

Ramp (Pos. or Neg.): Frequency (Hz), Amplitude (Vp-p or Ap-p),

Offset (Vd-c or Ad-c).

Square: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c).
Level: Timing (Sec), Amplitude (Vd-c or Ad-c).
• Local: the Start/Stop phases, available for sine and triangle only, can be set or

edited by modifying the existing segment (see PAR. 3.3.9.7).

• Remote: the start/stop phases are global parameters, applicable to all sine/ trian-

gular waveform segments.

• The Start/Stop points are independent and can have any value between 0° and

360°.

TABLE 1-2. BOP GENERAL SPECIFICATIONS (Continued)

SPECIFICATION

RATING/DESCRIPTION