Page 10 BM 700 Installation and Operating Instructions

Configuration menu (version 5.00 / 5.01)

Function (Fct.)

Input range

Description

1.0 OPERATION

1.1 DISPLAY

1.1.1 FCT.DISP

identical with 3.2.1

1.1.2 UNIT.LENGTH

identical with 3.2.2

1.1.3 UNIT.CONV.

identical with 3.2.3

2.0 TEST

2.1 HARDWARE

2.1.1 MASTER

Master hardware test.

2.1.2 DISPLAY

Display hardware test.

2.1.3 STATUS

Status information for Service

2.2 CUR.OUTP.I

2.2.1 VALUE I

Value display

Display of actual value of the current
output.

2.2.2 TEST I

Select 2 mA/4 mA/6 mA/...
... 20 mA/22 mA

Output of selected value to the current
output. With safety query.

2.4 FIRMWARE

2.4.1 MASTER

Display

Display of master firmware version.

2.4.2 DISPLAY

Display

Display of display firmware version.

3.0 INSTALL.

3.1 BASIS.PARAM

3.1.1 TANKHEIGHT

Select unit m/cm/mm/
inch/Ft
Enter
0.50 ... 20.00 [m]

Enter tank height (see explanatory
notes).
The unit entered here is also used for
all other length entries.

3.1.2 BLOCKDIST

Enter
0.10 [m] ... tank height

Enter block distance = non-measurable
range below bottom edge of flange (see
explanatory notes).

3.1.3 ANTENNA

Select STANDARD
WAVE-STICK

Select antenna type. WAVE-STICK for
all Wave-Stick versions, except type
"SW" for stillwells. All other =
STANDARD.

3.1.4 ANT.EXTENS.

Enter 0.00 [m] ... tank height Enter length of antenna extension.

3.1.5 DIST.PIECE

Enter 0 ... 2000 [mm] Enter length of distance piece above

flange (high temperature version =
120

mm).

3.1.6 STILLWELL

Select NO / YES
If “YES“: enter 25 ... 200 [mm]

Selection: without or with still well.
With still well: enter inside diameter in
[mm] (compensates different wave
speeds in still wells)

3.1.7 REF.OFFSET

Enter -10.00...0...+10.00 [m] Reference offset is added to measured

distance values.

3.1.8 TB.OFFSET

Enter -100.00...0...+100.00 [m] Tank bottom offset is added to

measured level values.

3.2 DISPLAY

3.2.1 FCT.DISP

Select LEVEL
DISTANCE
CONVERSION

Select function of display
(value to be displayed).
(see also explanatory notes)

3.2.2 UNIT.LENGTH

Select m/cm/mm/
inch/Ft/
PERCENT/BARGRAPH

Select unit for length value to be
displayed (only for level and distance).

3.2.3 UNIT.CONV.

Select m3/l(Liter)/US Gal/
GB Gal/Ft3/bbl/PERCENT/
BARGRAPH/USER UNIT

Select unit for conversion value to be
displayed (“volume table“).
(see explanatory notes)

3.2.4 USER UNIT

Text entry
10 characters

Enter user-defined unit for the
conversion table.

3.2.5 ERROR MSG.

Select NO/YES Select whether error messages to be

shown in display.