BM 700 Installation and Operating Instructions Page 13

Empty-tank spectrum

To enable the BM 700 to identify and blank out interference signals, e.g. caused by
fixed and moving tank internals, the tank profile (empty-tank spectrum) needs to be
recorded once only prior to (initial) start-up. For recording, the tank should be
completely empty and all moving parts (e.g. agitators) switched on. If major inter-
ference through internals is not expected, recording of the empty-tank spectrum can
also be dispensed with, since the factory has already carried out and stored a partial
empty spectrum of the flange system.

Empty-tank spectrum recording via display
After selecting menu item Fct. 3.5.2, press key

. The display then shows whether

the empty spectrum is currently ON or OFF. Then press the

key if no change is to

be made, or use the

key to choose between the following options:

ON: the empty-tank spectrum is (again) switched on and taken into account for
measurements.

OFF: the empty-tank spectrum is not taken into account for measurements, but
remains stored in the BM 700 and can be switched on again at a later date.

RECORD: the existing empty-tank spectrum is to be deleted and a new one
recorded.

After selecting "RECORD": if other parameters had previously been changed, the
query "ACCEPT YES" is first made as to whether they are to be stored. In this case,
confirm by pressing

. To record, use the

key to select one of the following options:

MAX. VALUES: (only maximum values are taken into account when the empty-
tank spectrum is recorded; useful e.g. with “difficult“ agitators).

AVERAGE: (values are averaged; this setting can be used for most
applications).

After selecting with the

key, press the

key to select TOTAL or the

key to select

PARTIAL.

When TOTAL is selected, the empty-tank spectrum is recorded over the entire
range (tank height).

If the tank has not been fully drained, the empty-tank spectrum can also be
recorded up to a certain distance, in which case the menu item PARTIAL should
be selected. When this has been selected, a query takes place by way of the

key concerning the distance value up to which the empty-tank spectrum is to be
recorded. The tank area below the current filling level is then excluded from the
empty-tank recording. It is recommended to maintain a safety distance of 20 to
30 cm from the actual product distance.

Subsequently press key

to start recording the empty-tank spectrum. The display

starts with “1000“ and counts down to “0“. The sign WAIT flashes in the display.
READY appears after approx. 2 minutes. Then press key

five times to store the

recorded empty-tank spectrum, which is taken into account for measurements.??

Empty-tank spectrum recording via PC-CAT
Connect the BM 700 and press in the display mode of PC-CAT the key combination
Ctrl-L. The type of empty-tank spectrum can be selected by one of the following keys:

1: Max.Values 4: Max. Partial A: Break
2: Average 5: Avg. Partial