KYORITSU 1109S User Manual

Page 4

Advertising
background image

̶ 2 ̶

● Never use the instrument in an explosive atmosphere especially

when making current measurements.

# CAUTION

● Before making measurements check that the range selector

switch is at a proper range position. Make sure to remove the
tips of the test leads from the circuit under test when changing
the measuring range during measurements.

● Do not apply voltage to the current or resistance ranges. It may

result in a fuse blow or instrument damage.

● Make certain to set the range selector switch to the OFF position

after every use.

● It is recommended that the range selector switch should be set

to the 250mA DC position to protect the instrument against the
possible shock or vibration in transit.

Note:

Take good care not to make voltage measurements with

the range selector switch at the 250mA DC position. The
fuse may blow or instrument get damaged.

● Do not expose the instrument to the direct sun, extreme

temperature and humidity or dew fall for a long period. Also, care
must be taken not to give a shock to the instrument by dropping
or inadequate handling.

● Since the meter cover has been given an anti-static treatment

do not rub it strongly with a dry cloth. Where anti-static charges
are present on the meter window after it has been used for a
long period of time, causing the meter pointer to deflect in an
abnormal way, coat its surface with anti-static chemicals for
plastics or wipe it with a dilute solution of anti-static fluid as a
temporary measure.

● Understanding Some of the Basics in Electrical Testing Before

Using the Multimeter