KYORITSU 1109S User Manual

Page 5

Advertising
background image

̶ 3 ̶

● Auxiliary Units (Prefixes)

There are a number of measurement units used for multimeters.

Volt (V), ampere (A) and ohm (Ω) are most widely used as

measurement units to indicate electrical potential, current and

resistance. However, it is not always straitforward to handle these

units as they sometimes too large or too small for practical use or

calculation. Prefixes are therefore used as auxiliary units to simplify

the usage of such measurement units.

The following table shows some of the examples:

Auxiliary Unit

M

k

m

μ

n

P

Read

mega-

kilo-

milli-

micro-

nano-

pico-

Multiply

10

6

10

3

10

-3

10

-6

10

-9

10

-12

Example

2MΩ

2000kΩ

2kΩ

2000Ω

250mA

0.25A

50μA

0.05mA

200nF

0.2μF

1000pF

0.001μF

● Notes on Voltage & Current Measurements

It is important to understand the basic differences between current

and voltage measurements for proper use of multimeters.

Voltage measurements are designed to detect potential difference

between two points. Make certain that the multimeter is connected in

parallel with the circuit under test.

Current measurements are intended to monitor the consumption of

current in the circuit resulting from the application of voltage. Make

sure to connect the multimeter in series with the circuit under test.

Generally speaking, the internal resistance of a voltmeter should

preferably be larger, while that of an ammeter should be smaller. If

the multimeter is erroneously connected in parallel with a circuit for

current measurements, the likely result will be the flow of excessive

current and aubsequent damage to the fuses and other electronic

components. To avoid such a potential danger it is necessary to have

a good understanding of voltage and current measurements.

Advertising