Specifications, 1 standards, 2 insulation resistance ranges and accuracy – KYORITSU 6017 User Manual

Page 6

Advertising
background image

4

3. Specifications

3-1 Standards

Designed to meet:
IEC 61557-2: Insulation Tester
IEC 61557-5: Earth Resistance Tester
IEC 61010-1 over-voltage category CAT. e 600V, pollution degree 2

3-2 Insulation Resistance Ranges and Accuracy

Function

1000V/2000MΩ

1000V

0-2000MΩ

50MΩ

1000VDC+20%

– 0%

1-1.6mADC or less

(Measured with an ammeter with a voltage drop of 500mV or less)

1-1.2mADC or less

(Measured with an ammeter with a voltage drop of 500mV or less)

at 1MΩ

2-1000MΩ

± 5% of indicated value

Ranges other than the above range, 0 and

± 10% of indicated value

0.7% of scale length

0.1-50MΩ

0.05-20MΩ

0.02-10MΩ

at 0.5MΩ

at 0.25MΩ

at 0.125MΩ

500VDC+20%

– 0%

250VDC+20%

– 0%

125VDC+20%

– 0%

2MΩ

1MΩ

0.5MΩ

0-100MΩ

0-50MΩ

0-20MΩ

500V

250V

125V

500V/100MΩ

250V/50MΩ

125V/20MΩ

Rated Test Voltage

Measuring Range

Mid-scale Value

Output Voltage at Open
Circuit

Short-circuit Current

Rated Test Current

Accuracy in Primary Effective
Measuring Range

Accuracy in Secondary
Effective Measuring Ranges

Accuracy at 0 and ∞

3-3 Earth Resistance Ranges and Accuracy: Precision Measurement

Function

Measuring Range

Accuracy

Output Current

X 1 Ω

0-12Ω

± 3% of full scale value

(The earth resistance at auxiliary earth spike is

within 10kΩ)

2.5mA AC max., 820 ±10Hz

0-120Ω

0-1200Ω

X 10 Ω

X 100 Ω