5 ac voltage ranges and accuracy, 6 input impedance, 7 battery life – KYORITSU 6017 User Manual

Page 7

Advertising
background image

5

3-4 Earth Resistance Ranges and Accuracy: Simplified Measurement

Function

Measuring Range

Accuracy

Output current is the same as in precision measurement.

X 1 Ω

0-12 Ω

0-120 Ω

0-1200 Ω

X 10 Ω

X 100 Ω

± 3% of full scale value

(Value subtracted test lead resistance)

3-5 AC Voltage Ranges and Accuracy

Function

Measuring Range

Frequency

Accuracy

600V AC

0-600V

50/60Hz

± 3% of full scale value

Earth Voltage: 60V AC

0-60V

3-6 Input Impedance

Function

Input Impedance

2.8MΩ

135kΩ

600V AC

Earth Voltage: 60V AC

3-7 Battery Life

Number of measurements within effective battery voltage range.

(5-second measurement /25-second pause cycle)

Function

Test Resistance

Number of Measurements

More than 1000 times

More than 2000 times

More than 3000 times

More than 3500 times

More than 3000 times

More than 4000 times

1MΩ

0.5MΩ

0.25MΩ

0.125MΩ

1200Ω

1200Ω

Earth ( 3 Pole )

Earth ( 2 Pole )

X 100 Ω

X 100 Ω

125V

250V

500V

1000V

Insulation Resistance

Advertising
This manual is related to the following products: