AC voltage measurement (Mains disconnection check)
The AC voltage measurement function is available on any ranges.

DANGER

¡

Never make measurement on a circuit in which the electrical
potential exceeds AC/DC600V in order to avoid possible
shock hazard.

(Refer to Section 3. "Specification, AC voltage measurement.)

¡

When testing installation that has a large current capacity,

such as a power line, be sure to make measurement on the
secondary side of a circuit breaker in order to avoid personal
injury.

¡

Do not press the Test button or Remote control switch during
voltage measurement.

¡

Never short live conductors with the tip of a probe to avoid
personal injury.

¡

Do not make measurement with the Battery cover removed.

<Simplified measurement of DC voltage>

This instrument also provides simplified measurement of DC
voltage. Multiplying 0.9 by the indicated value at AC voltage scale
provides the DC voltage value. However, the polarity isn't
indicated.
DC voltage (V) = Indicated value at AC voltage scale x 0.9 (V)

CAUTION

Never press
test button

CAUTION

Never press
remote switch