― 16 ―

6−5 Using Guard Terminal (only for MODEL3315)

Shown below is an example of the insulation resistance
measurement of an electric wire. If Line probe is simply
connected to the wire conductor and Earth probe to the woven
metal shield as shown, a measurement error will be introduced
as this results in the measurement of the combined resistance
of insulation resistance and the surface leakage resistance at
the cut end of the electric wire. In order to remove this surface
leakage current, wind a guard wire around the cut end of the
conductor and connect it to the guard terminal of the instrument
using Guard Connection lead. Then, the surface leakage
current will bypass the meter of the instrument.

6−6 Backlight function

To facilitate working in dimly lit
situations, a backlight function
is provided which illuminates
the scale plate. Press the
backlight switch to operate this
function. The backlight will light
up for about 60 sec., and then
turned off automatically.

Meter

Voltage
Generator

Leakage Current

Core Conductor

Insulation

Guard Wire

Sheated
Cable

Insulation Resistance Tester

Advertising