KYORITSU 3316 User Manual

Page 7

Advertising
background image

― 5 ―

<Operating error>
Operating error (B) is an error obtained under the nominal operating
conditions, and calculated with the intrinsic error (A), which is an
error of the instrument used, and the error (En) due to variations.
According to IEC61557, the maximum operating error should be
within +/-30%.

B =︱A︱+ 1.15 ×

( E

12

+ E

22

+ E

32

A : Intrinsic error (%)
B : Operating error (%)

1

: Variation due to changing the temperature (%)

2

: Variation due to changing the Battery voltage (%)

3

: Variation due to changing the position (%)

Nominal operating conditions

Ambient temperature : 0 ∼ 40℃
Relative humidity

: 90% or less

External magnetic field : 400A/m or less
Position

: Horizontal ∼ ±90°

Battery voltage

: within "BATTERY. GOOD" range

< AC voltage range>

Measured voltage

Accuracy

0 ∼ 600V
within ±3% of max scale value (ACV range)
within ±5% of max scale value (Others)

Range

50V/10MΩ

125V/20MΩ

250V/50MΩ

500V/100MΩ

1000V/2000MΩ

0.05MΩ

0.125MΩ

0.25MΩ

0.5MΩ

1MΩ

at least 3500 times

at least 2500 times

at least 2000 times

at least 1000 times

at least 300 times

Possible number
of measurement

<Number of measurement>

Possible number of measurement within the "BATTERY.GOOD"
range.(Measure 5 sec., and take pause for 25 sec..)

Resistor

for test