Tc-4000 specification – Techne TC-4000 User Manual

Page 21

Advertising
background image

TC-4000 SPECIFICATION

Temperature

Temperature range

4°C to 99°C

Temperature set point precision

0.1°C

Block uniformity (over full range)

± 0.6°C

Temperature accuracy

± 1.0°C

End of program cooling below ambient

Yes

Heating/Cooling Rate

Heating rate, between 55°C and 90°C on a 50°C to 95°C step:

96 x 0.2 ml micro tubes

2.6°C/sec

60 x 0.5 ml micro tubes

2.5°C/sec

Cooling rate, between 90°C and 55°C on a 95°C to 50°C step:

96 x 0.2 ml micro tubes

1.3°C/sec

60 x 0.5 ml micro tubes

1.6°C/sec

The programmable ramp rate is settable to a maximum of 2.0°C/sec. Any value above 2.0°C/sec will
default back to 2.0 unless set for max. If the unit is set to max, the heat up rate will be 2.6°C/sec.

Heated Lid

Heated lid enable/disable

Yes

Selectable heated lid temperature

100°C to 115°C

Over-temperature cut-out

Fixed at 145°C

Temperature sensor

Thermistor

Heater type

Wire wound

Heater power

132W

Warm up time

3 min approximately
(ambient to 105°C)

The heated lid will only come on if the set temperature is above 35°C

Programming

Program naming using alpha keys

Yes

Program password protection

Yes

Number of programs (3 step)

80

Maximum number of steps per stage

35

Maximum number of stages per program

25

Minimum hold time

1 sec

Maximum hold time

18 hrs

Incremental/decremental temperature

Yes

Incremental/decremental step hold time

Yes

Programmable ramp rate (heat or cool)

Variable in 0.1°C/sec steps

Programmable ramp rate resolution

0.1°C

19