Prime elite specification – Techne Prime Elite User Manual

Page 18

Advertising
background image

16

PRIME ELITE SPECIFICATION

TEMPERATURE

Block temperature range

4°C or 20ºC below ambient to 100ºC

Block uniformity

<±0.3ºC

Temperature accuracy

<±0.2ºC

Gradient availability

10°C to 100°C

Maximum gradient

30ºC (16ºC with the 384-well block)

Minimum gradient

1ºC

Column uniformity with a 10ºC gradient

<±0.3ºC

Pre-run sample cooling

Yes, 4ºC

Temperature set point adjustment

0.1ºC

HEATING/COOLING RATE

Maximum heating rate

5.0ºC/s

Minimum heating rate

0.1ºC/s

Maximum cooling rate

3.0ºC/s

Minimum cooling rate

0.1ºC/s

HEATED LID

Selectable heated lid temperature

35 to 115ºC or off

Pre-heat lid

Yes

Warm up time from ambient

<2 min

Over-temperature cut-out

Yes

Regulated heated lid pressure

Automatic

The heated lid is only operational if the block temperature is set above 35ºC

PROGRAMMING*

Program interface

5.7” colour VGA LCD touch screen

Maximum number of programs stored

1000

Maximum number of stages per program

99

Maximum number of steps per stage

99

Maximum number of cycles per stage

99

Programmable ramp rate

Yes, 0.1ºC/s steps

Maximum hold time

99h 59m 59s

Minimum hold time

1s

Gradient calculator

Yes

Incremented/decremented temperature

Yes

Incremented/decremented time

Yes

Pause facility

Yes

Program naming

Alpha numeric plus symbols

Password protection

Yes

Keyword selection

Yes

Run completion time

Yes

End of program alarm

Yes

Auto resume on power failure

Yes with options of 10min, 30min, 1h or always

Oligonucleotide T

m

calculator

Yes, based on the Nearest-Neighbour method

1

*Programming for Satellite units is via a Prime Elite unit or

Techne Workbench PC software.

1

Breslauer, K.J.; Frank, R.; Blocker, H. and Marky, L.A. (1986) Proc. Natl. Acad. Sci. USA 83, pp

3746-3750.