10.4 Ion setup mode (pH 2100 meter only)

10.4.1

Entering ion setup mode


1. Turn

the

meter

ON. If the meter is not in ion measurement

mode, press EXIT and use >/CAL key to select ion. The mode
selected will flash. Press ENTER to confirm selection. The
meter is now in Ion measurement mode.


2. Press

SETUP to enter the setup mode. The display will flash

Setup Mode Activated, Please wait... See Figure 129.


3. Use

>/CAL key to select ion. See Figure 130. The mode

selected will flash. Press ENTER to enter Ion setup mode. You
are now in the Ion setup mode.


4. The first setup program in Ion mode is Ion Unit Setup: P3.0.

See Figure 131. If you want to skip to another program, use
the

/MI and ∇/MR keys to scroll to the desired program.


5. To exit the setup mode at anytime, press EXIT. The meter

returns to the ion measurement mode.

NOTE: To exit the setup mode at anytime without confirming
changes, press EXIT. If you have made any changes to any
programs and press EXIT, these changes will be confirmed. To
reset meter to default factory settings, see setup program Meter
Reset P4.7
(page 54).



10.4.2

Ion unit setup mode P3.0


Use this program to select Ion measurement units: ppm, ppt, mg/L,
g/L, mmol/L, or mol/L

1. Enter ion setup mode as described in above in Section 10.4.1

Entering ion setup mode. The first setup mode is Ion Unit
Setup: P3.0
See Figure 132.


2. Press

ENTER to enter program Ion Unit Setup: P3.0. Press

>/CAL key to toggle between ppm, ppt, mg/l, g/l, mmol/l or
mol/l. The unit selected will flash. See Figure 133.


3. Press

ENTER to confirm selection and return to the setup

mode. To continue to the next setup program, press

/MI. To

continue to the previous program press

/MR. Press EXIT to

return to the ion measurement mode.

Setup: Ion Meter
Select & Press ENTER

Setup Mode Activated

Please wait. . .

Figure 129

Figure 130

Figure 131

Ion Unit Setup : P3.0
Press ENT. to Select

ppm ppt mg/1

g/l mmol/l mol/l

Ion Unit Setup :P3.0

Press ENT. to Select

Figure 132

Figure 133