17

1077 Technical Manual

P a g e

| 17

Set the 1077 to Load function with 10mA range and digit switch set to 9999. Connect the
24V supply to the 1077 through the DMM. Check the 1077 takes 10mA from the 24V
supply.

C. Linearity

The linearity of the output is not adjustable and is determined by the precision resistors
mounted on the digit switch. These have been selected and adjusted in the factory and
should not require further alteration.

D. LCD Current Meter F.S.D

1. Set the 1077 to line mode and take

20mA from the output.

2. Adjust VR1 on LCD meter for

20,00 milliamps.

Advertising