Time Electronics 1077 Milliamp Transducer Simulator User Manual

Page 7

Advertising
background image

7

1077 Technical Manual

P a g e

| 7

1. Output Terminals

The output current is available on the two front panel terminals which are suitable for
normal wire compression or 4mm plug insertion.

2. Case Terminal

Connected only to the instrument case and is isolated from the circuitry. The case
provides an overall electrostatic screen and can be earthed as required to improve
rejection of noise pick-up.

3. Output On-Off Switch

Disconnects the output terminals from the 1077 circuitry. The 'Off' position provides
an open circuit on the output terminals.

4. Range and Supply On-Off Switch

Four press button switches with interlocking on the 3 range buttons to ensure that
only one range can be selected at one time.

5. Digit Switch

Four decade thumbwheel switches with 0-9 positions on each decade.

6. Voltage Limit Indicator

An L.E.D. display which indicates when the 1077 has insufficient voltage across its
output terminals to sink or drive the set current.
Important Note:
Voltages in excess of 40 volts across the output terminals will cause an
overload condition in the 1077 and the output fuse will blow.

7. Battery Level Indicator

Located on the side of the instrument and continuously monitors the battery voltage.
A minimum mark indicates when the batteries need recharging.

8. Re-charger Socket

Also located on the side of the instrument. The mains re-charger is a separate unit
and the output is via a flying lead with a non-reversible plug. Charge current is 50
mA and recharge time is 12-14 hours.

9. Load/Line/Source Switch

A three position switch which selects the function of the 1077.

10. LCD Meter

Displays the output current taken when operating in 'Line' mode (in Load/Source
mode displays Zero).

11. Output Voltage Adjustment

Sets maximum open circuit voltage in 'Source' mode and output voltage in 'Line'
mode (normally set for 24V output).