6 calibration level 21, Point correction 21, Calibration level 21 - 23 – West Control Solutions TB 40-1 Temperature Monitor User Manual

Page 21: Offset correction 21, Rм щ, R ù r, Rr ù r, Rr ù, Rr щ м и

Advertising
background image

Offset correction (

ConF/ InP.1 / Corr =1 ):

InL.1: The input value of the scaling point is displayed.

The operator must wait, until the process is at rest.
Subsequently, the operator acknowledges the input value by pressing
key Ù.

OuL.1: The display value of the scaling point is displayed.

Before calibration,

OuL.1 is equal to InL.1.

The operator can correct the display value by pressing keys ÈÌ .
Subsequently, he confirms the display value by pressing key Ù.

Calibration level

TB40-1

21

r

М

Щ

3 sec.

r

PArA

CAL

r Ù r

InP.1

Ù

r

InL.1

r

r Ù

r

:

OuL.1

r

r Ù

End

r

r Щ

М

И

125

126

°C

Err