6.

"DSP HI" (display high) enter in display value

corresponding to high input value.

b.

Method 2 (LD CAL)

NOTE: Input Signal must be applied to the unit for low and high

display values. The meter compares the input signal to the

display value entered and scales the meter based on the two

inputs.

1.

Raw input is displayed. This can be used to make sure

the correct input is being applied and/or that the meter is

properly measuring it. Push enter to advance.

2.

"DEC PT" selects decimal point. Push

SELECT key for

displaying decimal point selections and

ENTER key to

confirm choice.

3.

"DSP LO" (display low).

Either select "YES" or "NO" here.

"YES" allows you to apply corresponding low input

signal to input and enter in corresponding low display

value. Push

ENTER key to make displayed value equal

input. If “NO” is selected it skips over this program and

leaves values to what they were.

4.

"DSP HI" (display high).

Either select "YES" or "NO"

If "YES" is selected, this allows you to apply

corresponding high input signal to input and enter in

corresponding high display value. Push

ENTER key to

make displayed value equal input. If “NO” is selected it

skips over this program and leaves values to what they

were.

5.

"ENABLE" (Enable). This enables the above calibration

setup by the inputs. This allows for the low value and

high values to be set at different times and then enabled

when both are complete.

"NO" leaves calibration unchanged

"YES" enables this calibration and overrides any

previous calibration by the USER or SHUNT type

calibration