Nonconductive materials intervening between the sensor and target have
little or no effect on system output. Because of this, environmental
contaminants, such as oil, dirt, humidity, and magnetic fields, have
virtually no effect on system performance. The function of a noncontact
displacement measuring system is to monitor the position of a target
relative to some reference plane. The magnitude of the analog output is
dependent upon the relative position of the target within the sensor’s
electromagnetic field—the farther the sensor is from the target, the higher
the analog output.

Reference

Plane

Reference

Plane

Target

Target

System Configuration

The figure below shows a block diagram of the system configuration.
Every system contains an electronics oscillator / demodulator and a
sensor.

± Vdc input

POWER

CONDITIONING

STAGE

OSCILLATOR

STAGE

RF

CONTROL

STAGE

DC

CONTROL

STAGE

CALIBRATION

CONTROLS

Zero
Gain
Linearity (Coarse or Fine)

SENSOR DRIVE

CONNECTORS

System V out

TO SENSOR