Symptoms and causes – Kaman KD-2300 User Manual

Page 42

Advertising
background image

KD-2300 Instruction Manual

Appendix C: Troubleshooting

••••

39

Problems with sensor mounting and nonrepeatable fixturing can be
sources of trouble:

Make sure your calibration fixture is sound and repeatable.

When you move the sensor from the calibration fixture to the
application fixture, maintain the offset.

Make every effort to provide a thermally stable environment to prevent
drift. (Kaman can provide temperature compensation, but your best
choice is to control the environment if possible).

Inappropriate target materials can also lead to difficulty using the
system:

When you calibrate, use actual target materials in the same sizes.

Whenever possible use the actual application environment or simulate it
as closely as possible.

Do not make any arbitrary changes in sensor/cable and
measurement module combinations

Symptoms and Causes

The following information is a Symptom/Solution cross reference. If after
reviewing the previous checklist of common problems you don’t remedy the
problem you are having, find the symptom you are experiencing below and
refer to the list of possible causes. Symptoms are marked with letters;
causes/solutions are marked with numbers. For example, if you are
experiencing the symptoms described in C, refer to solution 3.

SYMPTOMS

A. A saturated output of +/- 6 to 7 volts, with no change in target position and

with target in range. See 2 (open only) and 3,5,6.

B. Unchanging output voltage, anywhere from +1 to -2, with change in target

position. See 2 (short only) and 3, 5, 6.

C. Output drifts with no change in target position. See 3.

D. Random changes in output with no change in target position. See 1, 3, 4, 5, 7

(for multiple channel system) and 8.

E. Non-repeatability. See 3, 4, 6, 7, 8.

F. Unable to calibrate, or while calibrating you ‘run out’ of calibration controls.

See 1, 2, 3, 5, 6, 8.

G. System does not meet linearity specification. See 3.

Advertising