KD-2300 Instruction Manual

Appendix C: Troubleshooting

••••

41

5.CAUSE:

A malfunctioning voltmeter.
SOLUTION:

Check the system output voltage with a different voltmeter, or inject
a know voltage in the voltmeter in use to verify that it is operating
properly.

6.CAUSE:

Coarse linearity misadjusted.
SOLUTION:

Recalibrate the system.

7.CAUSE:

Beat note interference when using more than one sensor.
SOLUTION:

Ensure that the sensors are at least 3 diameters apart, or
Synchronize your sensor and electronics as described in
APPENDIX B.

8.CAUSE:

Ground loop.
This problem is usually the result of electrical ground surges from other
equipment.
SOLUTION:

1.In sensors with non-isolated coils (3U, 6U), isolate the sensors
from the ground.

2.Isolate the systems(s) from potential grounds so that the ground is
only through a desirable path.