3 specifications, 1 geomagnetic sensor characteristics, Specifications – PNI RM3100 Sensor Suite User Manual

Page 6: Geomagnetic sensor characteristics, Table 3-1: geomagnetic sensor performance

Advertising
background image

PNI Sensor Corporation

Doc 1017252 r03

RM3100 & RM2100 Sensor Suite User Manual

Page 5 of 45

3 Specifications

3.1 Geomagnetic Sensor Characteristics

Table 3-1: Geomagnetic Sensor Performance

1

Parameter

Cycle Counts

2

Units

50

100

200

Field Measurement Range

3

-800 to +800

T

Gain

20

38

75

LSB/ T

Sensitivity

50

26

13

nT

Noise

30

20

15

nT

Noise Density @ Max. Single-Axis Sample Rate

1.2

nT/ Hz

Repeatability over 200 T

15

8

8

nT

Hysteresis over 200 T

15

nT

Linearity over 200 T

0.5

%

Maximum Single-Axis Sample Rate
(divide by 3 for max. 3-axis sample rate)

1600

850

440

Hz

Single-Axis Average Current @ 24 Hz Sample

Rate (equivalent to 3-axis @ 8 Hz)

70

135

260

µA

Bias Resistor, R

b

121

External Timing Resistor for Clock, R

EXT

33

k

Circuit Oscillation Frequency

180

kHz

Operating Temperature

-40 to +85

C

Footnotes:

1. Performance values are typical. Performance specifications established with a supply voltage of

3.0 V, a bias resistor of 121

Ω, an external timing resistor of 33 kΩ, and with measurements taken

at room temperature. Other bias resistors, external timing resistors and operating voltages may
be used, but performance will differ from the values listed. Contact PNI for additional information.

2. The cycle count setting (eg. 50 ,100, and 200) is user-configurable and set in the Cycle Count

Registers. See Section 4.1 for a discussion on how the RM3100 works, and Section 5.1 for how
to set the Cycle Count Registers.

3. Field measurement range is defined as the monotonic region of the output characteristic curve.

Field measurement range can be extended using different bias resistors.