3d magic operation, Standard mode, Cycle count registers – PNI RM3000 Sensor Suites User Manual

Page 29: Ons 5.1, 5 3d magic operation – standard mode, 1 cycle count registers

Advertising
background image

RM3000 & RM2000 Sensor Suite User Manual r08

Page 29 of 41

5 3D MagIC Operation

– Standard Mode

Note: This section discusses how to operate the 3D MagIC in Standard Mode. For a description of
operation in Legacy Mode, see Section 6.
The 3D MagIC operates in Standard Mode when pin #12 is

held LOW (grounded to DVSS).

The basic functions to be performed when operating the 3D MagIC are:

Setting the values in the Cycle Count Registers, and

Taking sensor measurements.

The user should first establish the number of cycle counts to be measured for each sensor by

writing to the Cycle Count Registers. This is followed by sending a command or series of

commands to make the sensor measurements. Assuming the user will use the same number of

cycle counts for subsequent measurements, it is not necessary to rewrite to the Cycle Count

Registers for subsequent sensor measurements.

In Standard Mode, the 3D MagIC provides two methods to take sensor measurements, which are

discussed later in this section:

Single-axis measurement (SAM), and

Multi-axis measurement (MAM).

The SAM Command Byte initiates a measurement for one sensor and sets up the 3D MagIC to

write the measured values out on the MISO line. The MAM Command Byte initiates a sensor

measurement for up to 3 sensors, and a later Command Byte sets up the 3D MagIC to write the

measured values out on the MISO line for up to 3 sensors. For two or three axis systems,

normally it is more efficient to operate using the MAM Command Byte.

5.1 Cycle Count Registers

Prior to sending a command to take a sensor measurement, it is necessary to write values to

the Cycle Count Registers. (The default value for the Cycle Count Registers is 512D, but

this was chosen for legacy reasons and is an inefficient value otherwise.) The Cycle Count

Registers establish the number of sensor oscillation cycles that will be counted for each

sensor in both the forward and reverse bias directions during a measurement sequence.

Increasing the cycle count value increases measurement resolution, but system noise limits

the useable resolution such that the maximum efficient cycle count value generally is around

200-300 cycle counts. Lowering the cycle count value reduces acquisition time, which

increases maximum achievable sample rate or, with a fixed sample rate, decreases power

Advertising
This manual is related to the following products: