2 hygromet : 3-wire version, E-m-hm-v1_12 – ROTRONIC HygroMet MP402H User Manual

Page 13

Advertising
background image

E-M-HM-V1_12

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroMet MP Humidity Temperature
Meteorological Probe: User Guide

Instruction Manual

Document Type

Page 13 of 20

Document title

© 2009-2011; Rotronic AG E-M-HM-V1_12

6.3.2

HygroMet : 3-wire version


Electrical diagram for voltage outputs

The maximum permissible cable length can be
determined under consideration of the voltage drop
caused by the current flowing to the devices
connected to the unit. The voltage drop in the cable
depends both on cable resistance and on the
equivalent resistance of the devices connected in
parallel to the unit. The total resistance connected to
each unit output should be at least 1000 ohms.
Cable resistance should not be more than 1/1000 of

the load resistance.

Electrical diagram for current outputs



The maximum permissible cable length, connecting
the unit to other devices, is determined by the total
resistance resulting from the addition of the cable
resistance and that of the devices connected in series
with the unit. This resistance should not exceed 500
ohms.



Wiring color code






















Wire color

Description

Gray

Power ground

Green

Power supply (+) 5…24 VDC (depends on output signal type)

White – labeled 1

Output 1 (+) – default: relative humidity or calculated
parameter

White – labeled 2

Output 1 (-) – A-GND 1

Brown

Output 2 (+) – default: temperature

Pink

Output 2 (-) – A-GND 2

Red

RS485 (RXD)

Blue

RS485 (TXD)

Black

Optional: additional Pt100 RTD probe – PT S

Purple

Optional: additional Pt100 RTD probe – PT AS

Orange

Optional: additional Pt100 RTD probe – PT R

Yellow

Optional: additional Pt100 RTD probe – PT AR

Cable shield

Drain – should be connected to power supply (-) or to an
earth ground