10 technical data, 1 specifications, Technical data – ROTRONIC HygroMet MP402H User Manual

Page 17: Specifications, E-m-hm-v1_12

Advertising
background image

E-M-HM-V1_12

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroMet MP Humidity Temperature
Meteorological Probe: User Guide

Instruction Manual

Document Type

Page 17 of 20

Document title

© 2009-2011; Rotronic AG E-M-HM-V1_12

10 Technical data

10.1 Specifications

General

HM 2-wire

HM 3-wire

Device type

Humidity temperature probe with
analog output signals

Humidity temperature probe with analog
output signals and RS485 interface

Circuit type

2-wire, loop powered

3-wire

Power supply and connections

HM 2-wire

HM 3-wire

Supply voltage

10…28VDC
V min = 10 V + (0.02 x Load*)
*Load resistance in ohms.

15…24 VDC
0…1 V outputs: 5…24 VDC
0…5 V outputs: 10…24 VDC

Nominal current consumption

2 x 20 mA

< 50 mA

Electrical connections

3 m cable with tinned ends

Polarity protection

Protective diode on V+


Humidity and temperature
measurement

See document E-M-HC2 Probes > Specifications


Calculated parameters

HM 2-wire

HM 3-wire

Psychrometric calculations

Dew point (Dp) above and below freezing
Frost point (Fp) below freezing and dew point above freezing
Wet bulb temperature (Tw)
Enthalpy (H)
Vapor concentration (Dv)
Specific humidity (Q)
Mixing ratio by weight (R)
Vapor concentration at saturation (Dvs)
Vapor partial pressure (E)
Vapor saturation pressure (Ew)


Start-up time and data refresh rate

HM 2-wire

HM 3-wire

Start-up time

14 s (typical)

3 s (typical)

Data refresh rate

5 s (typical)

1 s (typical)


Advertising