3

General description

3.1 Power supply

Depending on the circuit type, the HygroMet MP requires the following power supply:

a) 2-wire, loop powered version: 10…28 VDC - depending on the load connected to the output(s). The

minimum supply voltage can be determined as follows:

V min = 10 V + (0.02 x Load*) *Load resistance in ohms.

For the maximum load of 500

Ω, the minimum supply voltage is 10 + (0.02 x 500) = 20 VDC. With both

output circuits closed, the maximum current consumption is 40 mA.

b) 3-wire version: 15 to 24 VDC. With both output circuits closed, the maximum current consumption is 50

mA.

Depending on the type of analog output signal, the minimum supply voltage can be reduced as follows:

0…1 V outputs: 5 VDC minimum
0…5 V outputs: 10 VDC minimum

3.2 Measured parameters

The HygroClip HC2-S3 probe used with the HygroMet MP measures relative humidity with a ROTRONIC
Hygromer

®

IN1 capacitive sensor and temperature with a Pt100 RTD.

3.3 Calculated parameters

Using the ROTRONIC HW4 software, the HygroMet MP can be configured by the user to calculate one of the
following parameters:

o

Dew point (Dp) above and below freezing

o

Frost point (Fp) below freezing and dew point above freezing

o

Wet bulb temperature (Tw)

o

Enthalpy (H)

o

Vapor concentration (Dv)

o

Specific humidity (Q)

o

Mixing ratio by weight (R)

o

Vapor concentration at saturation (Dvs)

o

Vapor partial pressure (E)

o

Vapor saturation pressure (Ew)


Note: some of the above parameters depend on the value of the barometric pressure. Using the ROTRONIC
HW4 software, a fixed barometric pressure value can be specified. For instructions see the following HW4
manual: E-M-HW4v3-F2-014