Unit

HygroMet MP Humidity Temperature
Meteorological Probe: User Guide

Instruction Manual

Document Type

Page 9 of 20

Document title

© 2009-2011; Rotronic AG E-M-HM-V1_12

Functions

Applicability

Factory default

Humidity / temperature adjustment

3-wire

Device write protection

2-wire, 3-wire

Disabled (HygroMet and probe)

Limit humidity output to 100 %RH

2-wire, 3-wire

Enabled (probe)

Out-of-limit value digital / display alarm

2-wire, 3-wire

Disabled (HygroMet and probe)

Data recording (probe)

3-wire

Enabled (loop mode – 10 min. interval)

Automatic humidity sensor test

3-wire

Disabled (probe)

Humidity sensor drift compensation

3-wire

Disabled (probe)

Fail safe mode

2-wire, 3-wire

Disabled (probe)

Monitor probe alarms

2-wire, 3-wire

Enabled

Loss of communication with probe

2-wire, 3-wire

Disabled (HygroMet)

Simulator mode

2-wire, 3-wire

Disabled (HygroMet and probe)


o

For a detailed description of all AirChip 3000 main functions see document E-T-AC3000-DF-V1

o

Instructions regarding the configuration of the HygroMet MP and probe as well as access to the functions
are provided in the following manuals:

E-M-HW4v3-Main
E-M-HW4v3-F2-014
E-M-HW4v3-F2-001
E-M-HW4v3-DR-001
E-M-HW4v3-A2-001
E-M-AC3000-CP

o

The factory default setting for dew / frost point calculation is frost point below freezing

4.4 Interaction between the HygroMet MP and probe functions


It is important to note that when used together, the HygroMet MP transmitter and HC2 probe (HygroClip 2)
constitute a 2-component system. Each system component has its own microprocessor, firmware and
functions. Some of these functions are unique to each system component. Other functions are found in both
components.

2-wire, loop powered HygroMet version: due to the necessity of limiting the current consumption of the
combination of HygroMet and HC2 probe to less than 4 mA, several probe functions such as RH sensor test,
data recording and probe adjustment are not available.

The functions and settings of the HygroMet MP (HM) and HygroClip 2 probe (HC2) operate together as
indicated below:

Function / Setting

HM

HC2

Notes

Device protection

X

X

Individual to the HygroMet and HC2 probe

RS-485 address

X

X

Individual to the HygroMet and HC2 probe

Device Name

X

X

User defined description
The device name of the HC2 probe is not displayed by HW4 and is
replaced with the HygroMet Input Name

Calculation

X

X

Psychrometric calculation
HygroMet setting overrides HC2 probe setting