Dew point accuracy, E-m-hf7-v1_24, 2 dew point accuracy – ROTRONIC HF7 User Manual

Page 22

Advertising
background image

E-M-HF7-V1_24

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF7 Humidity Temperature
Transmitters: User Guide

Instruction Manual

Document Type

Page 22 of 23

Document title

© 2009-2011; Rotronic AG E-M-HF7-V1_24

Conformity with standards

HF72

HF73

CE / EMC immunity

EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11

Solder type

Lead free (RoHS directive)

Fire protection class

Corresponds to UL94-HB

FDA / GAMP directives

compatible

Environmental limits

HF72

HF73

Operating limits at electronics

-5

0 … +85 °C, 0…100 %RH, non condensing – valid for survival only

Temperature limits at probe

-

100…+150 °C (valid for type N, D and A)

Maximum humidity at sensor

100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
45 %RH at 125 °C (260 °F)
15 %RH at 150 °C (302 °F)

Maximum air velocity at probe

40 m/s (7,870 ft /min)

– depends on dust filter type

Critical environments

Humidity sensor: as per DV04-14.0803.02 - Critical chemicals

10.2 Dew point accuracy

The HF7 can be configured to calculate either the dew point or frost point based on the measurement of
relative humidity and temperature. The accuracy of this conversion varies, depending on the humidity and
temperature conditions as shown in the graph below:

Example: at a temperature of 2

0 ˚C, a frost point value of -25 ˚C is measured with an accuracy of

± 0.5

˚C or better.

Advertising