Factory default settings, E-m-hf7-v1_24, 2 factory default settings – ROTRONIC HF7 User Manual

Page 8

Advertising
background image

E-M-HF7-V1_24

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF7 Humidity Temperature
Transmitters: User Guide

Instruction Manual

Document Type

Page 8 of 23

Document title

© 2009-2011; Rotronic AG E-M-HF7-V1_24

MEASUREMENT LOOP VALIDATION

AirChip 3000 Functions

Description

► Simulator mode

Used to make the HF7 generate fixed values for the humidity,
temperature and calculated parameter. Can be configured,
enabled or disabled

DEVICE SAFEGUARDS

AirChip 3000 Functions

Description

► Device write protection

Used to protect the HF7 with a password to prevent
unauthorized digital access by a digital user. Can be configured,
enabled or disabled

PROCESS PROTECTION / PROTECTION OF OTHER DEVICES

AirChip 3000 Functions

Description

► Limit humidity output to 100 %RH

Used to prevent the humidity signal from exceeding
100 %RH when condensation forms on the sensor. Can be
enabled or disabled

► Out-of-limit value alarm

Used to specify the normal range for humidity, temperature and
the calculated parameter depending on the user application.
Can be configured, enabled or disabled

Out-of-limit values trigger a digital alarm which can be also be
seen on the optional display

► Bad sensor alarm

Built-in function. Cannot be disabled

A bad humidity or temperature sensor triggers a digital alarm
which can be also be seen on the optional display

► Fail safe mode

Used to specify a "safe" fixed value for humidity and for
temperature in the event of a sensor failure. Can be configured,
enabled or disabled

4.2 Factory default settings


Notes:

o Configuration of the HF7 by the user and access to its functions requires a PC with the ROTRONIC HW4

software (version 2.1.1 or higher) installed. Service cable AC3006 or AC3009 is used to connect the HF7
service connector to a USB port of the PC (see Maintenance > Service Cable).


Configurable Settings

Applicability

Factory default

Unit system (Metric or English)

HF72, HF73

As per ordering code

Analog signal type (4…20 mA or other)

HF73

As per ordering code

Psychrometric calculation

HF72, HF73

As per ordering code

Output 1 parameter, scale and unit

HF72, HF73

As per ordering code (%RH or DP)

Output 2 parameter, scale and unit

HF72, HF73

Temperature, unit as per ordering code


Functions

Applicability

Factory default