Mechanical installation, General guidelines, E-m-hf7-v1_24 – ROTRONIC HF7 User Manual

Page 9: 5mechanical installation, 1 general guidelines

Advertising
background image

E-M-HF7-V1_24

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF7 Humidity Temperature
Transmitters: User Guide

Instruction Manual

Document Type

Page 9 of 23

Document title

© 2009-2011; Rotronic AG E-M-HF7-V1_24

Humidity / temperature adjustment

HF72, HF73

Device write protection

HF72, HF73

Disabled

Menu access from keypad

HF72, HF73

Enabled

Limit humidity output to 100 %RH

HF72, HF73

Enabled

Out-of-limit value digital / display alarm

HF72, HF73

Disabled

Data recording

HF72, HF73

Enabled (loop mode

– 10 min. interval)

Automatic humidity sensor test

HF72, HF73

Disabled

Humidity sensor drift compensation

HF72, HF73

Disabled

Fail safe mode

HF72, HF73

Disabled

Simulator mode

HF72, HF73

Disabled


o

For a detailed description of all AirChip 3000 / HF7 main functions see document E-T-AC3000-DF-V1

o

Instructions regarding the configuration of the HF7 and access to its functions are provided in the
following manuals:

E-M-HW4v3-Main
E-M-HW4v3-F2-007
E-M-HW4v3-DR-001
E-M-HW4v3-A2-001

o

The factory default setting for dew / frost point calculation is frost point below freezing

5

Mechanical installation

5.1 General guidelines

Relative humidity is extremely dependent on temperature. Proper measurement of relative humidity requires
that the probe and its sensors be at exactly the temperature of the environment to be measured. Because of
this, the location where you choose to install the probe can have a significant effect on the performance of the
instrument. The following guidelines should guarantee good instrument performance:

a) Select a representative location: install the probe where humidity, temperature and pressure

conditions are representative of the environment to be measured.

b) Provide good air movement at the probe: air velocity of at least 200 ft/ minute (1 meter/second)

facilitates adaptation of the probe to changing temperature.

c) Avoid the following: (1) Close proximity of the probe to a heating element, a cooling coil, a cold or

hot wall, direct exposure to sun rays, etc. (2) Close proximity of the probe to a steam injector,
humidifier, direct exposure to precipitation, etc. (3) Unstable pressure conditions resulting from
excessive air turbulence.

d) Immerse as much of the probe as possible in the environment to be measured.


HF7 type N with 120 mm probe length:
to avoid measurement errors caused by temperature
gradients fully immerse the probe barrel in the environment to be measured


Advertising