EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER
OPERATION MANUAL

Transmille Ltd.

Page 12

Connect the EA001A to the adapter interface on the calibrator front panel using the

supplied connection lead. Connect the Unit Under Test (UUT) to the connector

marked Input From Thermocouple (socket).

The thermocouple measurement option allows the user to measure thermocouple probes

of the types supported by the EA001A adapter.

Using the menus


On entering thermocouple simulation mode, type K will be the default function.

The display below will be shown indicating the currently selected range and output status :

Once in thermocouple mode, the menus available at the bottom of the screen will change

to reflect the thermocouple mode options - the available menu items are :

Selects the manual cold junction compensation mode

This allows cold junction value to be set to zero.

Selects the automatic cold junction compensation mode

This allows the calibrator to use the built in temperature sensor within the

adapter’s thermocouple plug to compensate for the cold junction measured

between the adapter plug and the UUT socket.

Thermocouple type

Cold junction
compensation type
Auto CJ = Automatic
Manual CJ = Manual

Temperature reading

Output status
(On / Standby)

Manual Cold junction soft key

Auto Cold junction soft key

Change thermocouple type soft key

MAN CJ

AUTO CJ