EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER
OPERATION MANUAL

Transmille Ltd.

Page 8

Using The Thermocouple Simulation Function

To use the thermocouple simulation function, press the

key and select the following :

Connecting the UUT


Connect the EA001A to the adapter interface on the calibrator front panel using the

supplied connection lead. Connect the Unit Under Test (UUT) to the connector

marked Output to Temperature Meter (fixed cable).

The thermocouple simulation option allows the user to enter a temperature in °C and set

the calibrator to the specific voltage output which corresponds to the thermocouple type

selected.