EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER
OPERATION MANUAL

Transmille Ltd.

Page 5

Main Features

Dedicated External Adapter For the Ultimate in Accuracy


Temperature gradients / thermal EMFs which can be introduced by internally mounted
thermocouple simulation methods are one of the greatest sources of error with
thermocouple temperature calibration.

For the ultimate in accuracy, Transmille has designed a dedicated external adapter. This
keeps the electronics required to generate the low level signals used for thermocouple
simulation as close as possible to the measuring input of the thermometer. This allows the
signal to be as free from electrical noise as possible, and also eliminates errors caused by
heat generated by surrounding electronics in the calibrator.

Direct Connection - No compensation cables required


Uses neutral (copper) thermocouple connectors

For direct connection to any thermocouple type input without the need for compensation
cables

CJC sensor

built in to
thermocouple
input socket for
automatic cold
junction
compensation

CJC sensor built in to thermocouple

plug for automatic cold junction
compensation