Ea002 – 2/10/50 turn coil adapter (option), Ea002, Dapter – Transmille 1000-Series User Manual

Page 58: Ption

Advertising
background image

1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 58

EA002 – 2/10/50 Turn Coil Adapter (Option)


The 1000 Calibrators can be used with the optional current coil EA002 (2, 10 and 50
turn coils). This enables currents both DC and AC of up to 500A to be produced for
the calibration of current clamp meters.
The current from the calibrator is connected to the appropriate coil connections. The
current output from the calibrator is then multiplied by the number of turns in that
coil, simulating a higher current for the clamp meter to measure.
















Complete the following procedure to select 500A DC using the 50 turn coil.