Simulated resistance commands, Imulated, Esistance – Transmille 1000-Series User Manual
Page 85: Ommands, Function mode activation command, Standby mode, Output
1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 85
Simulated Resistance Commands
Function
Mode Activation Command
Simulated Resistance
I2
Range
Command Output
0 Ohms – 10.0 Ohms
R26
Ohms
10 Ohms – 50 Ohms
R53
Ohms
50 Ohms – 100 Ohms
R27
Ohms
100 Ohms – 1k Ohms
R28
kOhms
1kOhms – 9.99kOhms
R29
kOhms
10kOhms – 99.9kOhms
R30
kOhms
100kOhms – 999kOhms
R31
MOhms
1MOhms – 9.99MOhms
R32
MOhms
Standby Mode
Standby ON
S1
Standby OFF
S0
Output
Set Output
x.xxxx
The resistance section consists of a set of range commands which are used in
conjunction with the standby command. To enable a resistance to be set, the
following command sequence should be used :
<FUNCTION>/<RANGE>/<OUTPUT>/<STANDBY CONDITION><CR>
Examples :
8 kOhms
5 MOhms
60 Ohms
I2/R29/O8/S0<CR>
I2/R33/O5/S0<CR>
I2/R53/O60/S0<CR>
I2 = Simulated Resistance Function
R29 = 1kOhms – 9.99kOhms Range
O8 = 8 kohms Output
S0 = Standby OFF (i.e. output ON)
<CR> = Carriage Return (ASCII 13)
I2 = Simulated Resistance Function
R32 = 1MOhms – 9.99 MOhms Range
O5 = 5 MOhms Output
S0 = Standby OFF (i.e. output ON)
<CR> = Carriage Return (ASCII 13)
I2 = Simulated Resistance Function
R53 = 50 Ohms – 99.9 Ohms Range
O60 = 60 Ohms Output
S0 = Standby OFF (i.e. output ON)
<CR> = Carriage Return (ASCII 13)