Simulated resistance commands, Imulated, Esistance – Transmille 1000-Series User Manual

Page 85: Ommands, Function mode activation command, Standby mode, Output

Advertising
background image

1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 85

Simulated Resistance Commands

Function

Mode Activation Command

Simulated Resistance

I2

Range

Command Output

0 Ohms – 10.0 Ohms

R26

Ohms

10 Ohms – 50 Ohms

R53

Ohms

50 Ohms – 100 Ohms

R27

Ohms

100 Ohms – 1k Ohms

R28

kOhms

1kOhms – 9.99kOhms

R29

kOhms

10kOhms – 99.9kOhms

R30

kOhms

100kOhms – 999kOhms

R31

MOhms

1MOhms – 9.99MOhms

R32

MOhms

Standby Mode

Standby ON

S1

Standby OFF

S0

Output

Set Output

x.xxxx


The resistance section consists of a set of range commands which are used in

conjunction with the standby command. To enable a resistance to be set, the

following command sequence should be used :

<FUNCTION>/<RANGE>/<OUTPUT>/<STANDBY CONDITION><CR>

Examples :

8 kOhms

5 MOhms

60 Ohms

I2/R29/O8/S0<CR>

I2/R33/O5/S0<CR>

I2/R53/O60/S0<CR>

I2 = Simulated Resistance Function

R29 = 1kOhms – 9.99kOhms Range

O8 = 8 kohms Output

S0 = Standby OFF (i.e. output ON)

<CR> = Carriage Return (ASCII 13)

I2 = Simulated Resistance Function

R32 = 1MOhms – 9.99 MOhms Range

O5 = 5 MOhms Output

S0 = Standby OFF (i.e. output ON)

<CR> = Carriage Return (ASCII 13)

I2 = Simulated Resistance Function

R53 = 50 Ohms – 99.9 Ohms Range

O60 = 60 Ohms Output

S0 = Standby OFF (i.e. output ON)

<CR> = Carriage Return (ASCII 13)

Advertising