Ac current commands, Urrent, Ommands – Transmille 1000-Series User Manual

Page 80: Function range command, Standby mode, Output, Ac frequency

Advertising
background image

1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 80

AC Current Commands

Function

Range

Command

AC Current

100uA

R17

1mA

R18

10mA

R19

100mA

R20

1A

R21

10A

R22

Standby Mode

Standby ON

S1

Standby OFF

S0

Output

Set Output

O (not zero)

AC Frequency

Fxxxxx

E.G. 10kHz = F10000

The AC current section consists of a set of range commands which are used in
conjunction with the standby and output command. To enable a AC current to be set
and an output assigned, the following command sequence should be
used:

<RANGE>/<OUTPUT>/<FREQUENCY>/<STANDBY CONDITION><CR>

For example, to get 10mA @ 1 kHz AC with the output switched on, the command
is:

R19/O10/F1000/S0<CR>

R19 = 10mA Range (as detailed in the table above)
O10 = 10mA Output
F1000 = 1 kHz Frequency
S0 = Standby OFF (i.e. output switched ON)
<CR> = Carriage Return (ASCII character 13)