UEi Test Instruments C75OILKIT User Manual

Page 12

Advertising
background image

EFF (G): Combustion efficiency calculation displayed in percentage.

Gross (G) or Net (N) can be set (see MENU). The calculation
is determined by the fuel type and uses the calculation in
British Standard BS845. The efficiency is displayed during a
combustion test, “- - - -” is displayed while in fresh air.

XA IR % : Excess air calculated from the measured oxygen and type of

fuel used. Displays reading during a combustion test. “- - - -” is
displayed while in fresh air.

A MB :

Boiler air INL ET t e m p e rature used to calculate the
NET t e m p e ra t u r e .

: Displays the approximate battery level as follow s :

Full battery level

B a t t e ry at 75 %

B a t t e ry at 50 %

B a t t e ry at 25 %

When the display flashes this, it indicates the batteries are at less than 10 %
of charge and should be replaced, readings may be affected if the analyzer
is used with low power batteries.

C75-MAN

P. 9

S p e c i f i c a t i o n s

*1

Using dry gases at STP

*2

Calculated

A p p e n d i c e s

Main Display Parameters

O 2 :

Oxygen reading in percentage (%).

T f :

Te m p e rature is measured by the flue gas probe in Centigrade
or Fahrenheit. Will show ambient temperature after fresh air
ca l i b ration and “- - - -” if the flue probe is disconnected.

C O :

Carbon Monoxide reading displayed in ppm (parts per
million). “- - - -” is displayed if there is a fault with the CO
sensor or the instrument has not set to zero correctly,
switch off instrument and try again.

C O a :

Carbon Monoxide air-free reading referenced to an oxygen
level of 0%. Do not confuse this reading with the actual CO
reading as detailed above.

C O

2

:

Carbon Dioxide calculation determined by the type of fuel.
This only shows a reading when a combination test is being
carried out. “- - - -” is displayed while in fresh air.

∆T :

Net temperature calculated by deducting the A MBIENT
(or INL ET) temperature from the measured FL UE
t e m p e rature. Displays in either Centigrade (˚C) or Fahrenheit
(˚F) and will display “- - - -” if the flue probe is not connected.

Parameter

Range

Resolution

Accuracy

Temp Measurement
Flue temperature

1.0˚ F/C

±5˚F (2.0˚C)

32 - 1112˚F

±0.3% reading

0 - 600˚C

Inlet temperature

1˚ F/C

±1˚ F/C

32 - 212˚F

±0.3% reading

0 - 100˚C

Temp (Nett)

*2

1.0˚ F/C

±5˚F (2˚C)

32 - 1112˚F

±0.3% reading

0 - 600˚C

Gas Measurement
Oxygen

0.1%

±0.2%

*1

0 - 21%

*Carbon Monoxide

1 ppm

±10 ppm <100 ppm

*1

0 - 1000 ppm

±5% reading

Carbon Dioxide

*2

0.1%

±0.3% reading

0 - 30%

Efficiency

*2

0.1%

±1.0% reading

0 - 99.9%

Excess Air

*2

0.1%

±0.2%

0 - 250%

Pre-programmed fuels

Natural gas, Light Oil, Propane, Butane, LPG

Dimensions
Weight

1 kg. / 2.2 lb.

Handset

200mm/7.9” x 45mm/1.8” x 90mm/3.5”

Probe

(L) 300mm/7.9” x (Dia) 6mm/0.25” with 200mm/7.8”
long stainless steel shaft, type K thermocouple and
3m/6ft long neoprene hose

Ambient operating

+32˚ - 104˚F (0˚ - 40˚C)
10% to 90% RH non-condensing

Battery life

4 “AA” cells

range

>8 hours using Alkaline “AA” cells

AC adapter (optional)

Input: 110 V AC
Output: 9 V DC regulated