Brewer Access High-Low User Manual

Page 18

Advertising
background image

Document # 100452 RevL

18

Printed in USA © 2013

TROUBLESHOOTING

Operating Test

It is possible to verify normal operation of the

series 6000 tables by performing the following

series of tests.

ALL MODELS

Lift test

1. Insure ON/OFF switch is in ON position.

2. Press and hold the TABLE UP pedal. The

table should rise to the full up position

(about 37”) and stop. There must be no

continuing motor hum after the table stops.

3. Release the TABLE UP pedal.

4. Press the TABLE DOWN pedal. While

the table is going down pressing on each

corner of the back of the seat must cause

the table to momentarily stop.

5. The table should go to the full down posi-

tion. Release the TABLE DOWN

pedal.

6. Reach under the front section bottom plate

and lift up momentarily on each corner of

the plate. This should activate the lift caus-

ing the table to rise while the plate is lifted.

7. Turn the ON/OFF switch, located on the

back of the rear cabinet, to the OFF posi-

tion. Verify that pressing any of the foot

pedals does not cause motion. Turn the

ON/OFF switch to the ON position.

MODELS 6500, 6501, 6800 and 6801 ONLY

Powered Back test

1. Press the TABLE UP pedal to raise the table

above 28” so the back may be lowered without

contacting the back cabinet.

2. Press the BACK DOWN pedal. The back

should go to the full recline position, which is

level.

3. Press the BACK UP pedal. The back should

go to the full up position.

4. Press the BACK DOWN pedal. While the

back is going down pressing on any corner of

the back of the seat must cause the back to

momentarily stop.

5. Press the blue RETURN TO CHAIR button on

the foot control. While the chair is in motion

momentarily press the red STOP button. The

chair must stop and remain stopped.

6. Press the blue RETURN TO CHAIR button on

the foot control. While the chair is in motion

momentarily press the BACK UP pedal. The

chair must stop and remain stopped.

7. Press the blue RETURN TO CHAIR button on

the foot control. While the chair is in motion

momentarily press the BACK DOWN pedal.

The chair must stop and remain stopped.

8. Press the blue RETURN TO CHAIR button

on the foot control. While the chair is in mo-

tion momentarily press the TABLE UP pedal.

The chair must stop and remain stopped.

9. Press the blue RETURN TO CHAIR button on

the foot control. While the chair is in motion

momentarily press the TABLE DOWN pedal.

The chair must stop and remain stopped.

10. Raise the table to the full up position, and then

lower the back so the table is in the high table

position. Press the blue RETURN TO CHAIR

button. The table should first raise the back

then lower to the chair low position.